FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2152237
·
Received June 23, 2011
Report
- Report Number
- 1720753-2011-08461
- Event Type
- Malfunction
- Date Received
- June 23, 2011
- Date of Event
- June 14, 2011
- Report Date
- June 23, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION AND REPLACED THE BATTERY PACKS. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM'S SCREENS WENT BLANK AND THE SYSTEM DISPLAYED A PRECHARGE VOLTAGE ERROR MESSAGE DURING A CASE. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |