FDA Adverse Event Malfunction Summary report: N

6800

MDR report key: 2152228 · Received June 23, 2011

Report

Report Number
1720753-2011-08440
Event Type
Malfunction
Date Received
June 23, 2011
Date of Event
June 16, 2011
Report Date
June 23, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REPRESENTATIVE CONDUCTED AN ON SITE INVESTIGATION. THE WORKSTATION PRINTED CIRCUIT BOARDS WERE RESEATED AND THE SOFTWARE WAS RELOADED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6800

Patients

Seq Age Sex Outcome Treatment
1