FDA Adverse Event Malfunction Summary report: N

PLUM 1.6 WITH DATAPO

MDR report key: 2152217 · Received June 23, 2011

Report

Report Number
9615050-2011-00448
Event Type
Malfunction
Date Received
June 23, 2011
Date of Event
May 25, 2011
Report Date
May 25, 2011
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K911401
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. A VISUAL INSPECTION FOUND THAT THE GROUND PRONG ON THE AC POWER CORD WAS BROKEN. THE PROBABLE CAUSE FOR THE BENT BLADE ON THE AC POWER CORD IS USE IN THE CUSTOMER ENVIRONMENT. IF THE AC POWER CORD IS ATTEMPTED TO BE REMOVED FROM AN OUTLET BY PULLING AT ANGLE, IT IS POSSIBLE TO DAMAGE THE AC POWER CORD BLADE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE GROUND PRONG ON THE AC POWER CORD PLUG WAS BROKEN. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLUM 1.6 WITH DATAPO 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA