FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2152141 · Received July 7, 2011

Report

Report Number
6000001-2011-10676
Event Type
Malfunction
Date Received
July 7, 2011
Date of Event
June 9, 2011
Report Date
June 13, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE 804:26 WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THIS CONDITION WAS ASSIGNED TO A DEFECTIVE PUMP HEAD MODULE. THE PUMP HEAD MODULE WAS REPLACED TO CORRECT THIS CONDITION. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 6.13.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE HAS NOT BEEN SERVICED PRIOR TO THIS EVENT FOR THE REPORTED CONDITION.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A COLLEAGUE INFUSION PUMP IN WHICH FAILURE CODE 804:26 OCCURRED. IT IS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1