HOOK MCO805 165MM 2.0MM 90DEG BLUNT
Report
- Report Number
- 3003249645-2025-00007
- Event Type
- Malfunction
- Date Received
- March 4, 2025
- Date of Event
- January 29, 2025
- Report Date
- March 4, 2025
- Manufacturer
- INTEGRA MICROFRANCE S.A.S.
- Product Code
- JYL
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHARMACIST
- Health Professional
- Yes
Narratives
THE 165MM 2.0MM 90DEG BLUNT HOOK (MCO805) WAS NOT RETURNED TO THE MANUFACTURER. LOT NUMBER INFORMATION WAS NOT PROVIDED; THUS, AN EVALUATION OF THE DEVICE COULD NOT BE PERFORMED, AND A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. IT WAS NOTED THAT IF THE END OF THE HOOK BROKE DURING THE INTERVENTION, IT IS PROBABLY THAT IT HAD BEEN WEAKENED BY AN IMPACT DURING CLEANING OR BY A FALL. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE IN THE FUTURE, THIS COMPLAINT WILL BE REOPENED, AND THE RESPECTIVE EVALUATION PERFORMED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR ISSUES. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.
IT WAS REPORTED THAT "IN THE ENT AND HEAD AND NECK SURGERY OPERATING ROOM, DURING A MENINGEAL PLASTY FOR EXCISION SURGERY OF A LEFT ETHMOID ADENOCARCINOMA, BREAKAGE OF THE END OF THE STAINLESS-STEEL BUTTONED HOOK"/165MM 2.0MM 90DEG BLUNT HOOK (MCO805) OCCURRED. "THE INSTRUMENT BELONGS TO THE OPERATING TRAY CALLED 'MENINGEAL TRAY'." IT WAS REPORTED THAT "THE SIZE OF THE BROKEN PART WAS ESTIMATED AT 2MM. THE FRAGMENT WAS NOT FOUND IN THE OPERATING FIELD. THE PATIENT WAS SUBJECTED TO CLINICAL MONITORING AND A CT (COMPUTED TOMOGRAPHY) SCAN THAT DID NOT REVEAL THE PRESENCE OF THE BROKEN ELEMENT." THERE WERE REPORTEDLY NO OTHER CLINICAL CONSEQUENCES FOR THE PATIENT NOR EXTENSION OF PROCEDURE TIME. THE INSTRUMENT WAS DISCARDED AND REPLACED IN THE OPERATING TRAY AFTER THE OPERATION. IT WAS REPORTED THAT THIS WAS AN ISOLATED INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2275377 | HOOK MCO805 165MM 2.0MM 90DEG BLUNT | PFM11 | JYL | INTEGRA MICROFRANCE S.A.S. | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |