COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-10641
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- June 26, 2011
- Report Date
- June 30, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). ADDITIONAL INFORMATION: THIS INVOLVED A REMEDIATED COLLEAGUE INFUSION PUMP WITH USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.90. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE HAS BEEN SERVICED SIX TIMES PRIOR TO THIS EVENT. THE DEVICE HAS BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION OF "DAMAGED BATTERY". "PREVIOUS SERVICE ON 9/25/2007, 2/22/2010, 7/20/2010 & 12/8/2010 WAS FOR "DAMAGED BATTERY", THE BATTERIES AND BATTERY HARNESS WERE REPLACED. DURING PREVIOUS SERVICE ON 4/13/2007 & 6/24/2006 THE BATTERIES AND BATTERY HARNESS WERE REPLACED. EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY ALARM WAS CONFIRMED BY BAXTER PERSONNEL DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS NOT IDENTIFIED. THE DEVICE WAS RETURNED UNREPAIRED. THERE WERE NO UPGRADES/REPAIRS PERFORMED ON THIS DEVICE. SERVICE WAS UNABLE TO VERIFY FUNCTIONALITY OF THE DEVICE DUE TO ESTIMATE REFUSAL BY CUSTOMER. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH DAMAGED BATTERY ALARM; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVED; THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SOFTWARE VERSION IS CURRENTLY UNKNOWN AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |