COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-10621
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- June 27, 2011
- Report Date
- June 28, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF BAD DISPLAY SCREEN GRAPHICS WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS NOT IDENTIFIED. NO REPAIRS HAVE BEEN PERFORMED DUE TO THE CUSTOMER?S REFUSAL OF ESTIMATE, NO FURTHER CORRECTION WAS MADE CONCERNING THIS EVENT AND THE PUMP WAS RETURNED UNREPAIRED. ADDITIONAL INFORMATION: THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 6.13.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE HAS NOT BEEN SERVICED PRIOR TO THIS EVENT FOR THE REPORTED CONDITION.
THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH AN LCD SCREEN THAT HAD BAD GRAPHICS; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS CONDITION WAS FOUND IN THE GENERAL PT WARD UPON POWER UP. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVED; THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SOFTWARE VERSION IS CURRENTLY UNKNOWN AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |