J-VAC RESERVOIR
Report
- Report Number
- 2210968-2011-00869
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- June 14, 2011
- Report Date
- June 14, 2011
- Manufacturer
- ETHICON
- Product Code
- KOG
- PMA / PMN Number
- K953655
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
(B)(4). RESULTS: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY EVALUATED. THE ANTI REFLUX VALVE WAS DETACHED. CONCLUSION: THE DEVICE WAS RECEIVED IN A CONDITION WHICH DOES NOT MEET SPECIFICATION.
(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE PRODUCT EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC RIGHT HEMICOLECTOMY ON (B)(6) 2011 AND A DRAIN WAS PLACED. ON (B)(6) 2011, IT WAS FOUND THAT THE ANTI-REFLUX VALVE WAS DETACHED AND FELL INTO THE RESERVOIR. ANOTHER LIKE WAS USED WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | J-VAC RESERVOIR | WOUND DRAINAGE SYSTEM | KOG | ETHICON | NA | J1022261 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |