FDA Adverse Event Malfunction Summary report: N

J-VAC RESERVOIR

MDR report key: 2151981 · Received July 7, 2011

Report

Report Number
2210968-2011-00869
Event Type
Malfunction
Date Received
July 7, 2011
Date of Event
June 14, 2011
Report Date
June 14, 2011
Manufacturer
ETHICON
Product Code
KOG
PMA / PMN Number
K953655
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY EVALUATED. THE ANTI REFLUX VALVE WAS DETACHED. CONCLUSION: THE DEVICE WAS RECEIVED IN A CONDITION WHICH DOES NOT MEET SPECIFICATION.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE PRODUCT EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC RIGHT HEMICOLECTOMY ON (B)(6) 2011 AND A DRAIN WAS PLACED. ON (B)(6) 2011, IT WAS FOUND THAT THE ANTI-REFLUX VALVE WAS DETACHED AND FELL INTO THE RESERVOIR. ANOTHER LIKE WAS USED WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 J-VAC RESERVOIR WOUND DRAINAGE SYSTEM KOG ETHICON NA J1022261

Patients

Seq Age Sex Outcome Treatment
1