FDA Adverse Event Malfunction Summary report: N

SINGLE PM KIT VAMP PLUS

MDR report key: 2151964 · Received July 7, 2011

Report

Report Number
2015691-2011-15838
Event Type
Malfunction
Date Received
July 7, 2011
Date of Event
June 13, 2011
Report Date
June 14, 2011
Manufacturer
EDWARDS LIFESCIENCES DR
Product Code
DXO
PMA / PMN Number
K925638
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONE DPT-VAMP KIT WAS RETURNED FOR ANALYSIS WITH PRESSURE AND IV TUBING ATTACHED. PRIMING SOLUTION WAS VISIBLE INSIDE OF THE KIT. THE DPT ZEROED AND SENSED PRESSURE ACCURATELY . PRESSURE READINGS WERE ALSO STABLE DURING TESTING. PRESSURE DID NOT SHOW ANY DRIFT DURING TESTING FOR 8 HOURS. ELECTRICAL TESTING SHOWED THAT BOTH INPUT AND OUTPUT IMPEDANCE MET SPECIFICATIONS. NO VISIBLE DEFECT WAS FOUND ON THE SUPERCOMAL CABLE CONNECTOR. PRESSURE TESTING WAS PERFORMED USING A GE MONITOR AND PRESSURE GAUGE. ELECTRICAL AND CONTINUITY TESTING WAS PERFORMED USING A DIGITAL MULTIMETER. VISUAL EXAMINATION WAS PERFORMED UNDER 10X MAGNIFICATION. THE COMPLAINT OF A FALSE PRESSURE READING WAS NOT CONFIRMED AND NO EVIDENCE OF A MANUFACTURING DEFECT WAS FOUND DURING THE ANALYSIS. IT COULD NOT BE DETERMINED IF ANY CLINICAL FACTORS MAY HAVE CONTRIBUTED TO THIS EVENT. NO ACTIONS WILL BE TAKEN AT THIS TIME. A DEVICE HISTORY RECORD REVIEW WAS COMPLETE AND DOCUMENTED THAT THE DEVICE MET ALL SPECIFICATIONS UPON DISTRIBUTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE GAVE A FALSE BLOOD PRESSURE READING OF 190/140. THE WAVEFORM WAS NORMAL AND THE NURSE WEANED OFF THE "PRESSURES" (BELIEVED TO MEAN VASOPRESSORS). THE RN INFORMED THE MD AND RECEIVED AN ORDER TO GO BY ARTERIAL LINE. THE CUFF READING WAS 70/40'S. THE CUSTOMER DID NOT FEEL COMFORTABLE WITH THE ARTERIAL LINE READING. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SINGLE PM KIT VAMP PLUS DISPOSABLE PRESSURE TRANSDUCER DXO EDWARDS LIFESCIENCES DR PXVP2284 59026450

Patients

Seq Age Sex Outcome Treatment
1