FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2151841 · Received July 6, 2011

Report

Report Number
2531779-2011-04723
Event Type
Malfunction
Date Received
July 6, 2011
Report Date
June 8, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION AND EVALUATED. CONFIRMED THE BOLUS BUTTON IS NOT RESPONDING TO BUTTON PRESSES. BOLUS BUTTON COVER WAS REMOVED TO INVESTIGATE AND IT WAS NOTED THAT THE INTERNAL PLUG IS MISSING. TESTING CONFIRMED ALL BUTTONS ON THE KEYPAD ARE RESPONSIVE TO BUTTON PRESSES AND ARE FUNCTIONAL. THE KEYPAD HAS NO VISIBLE SIGN OF DAMAGE. REMOVED KEYPAD COVER TO CHECK CONDITION OF THE BUTTON CONTACTS AND NO ISSUES FOUND.

Description of Event or Problem · 1

THE PATIENT CLAIMED THAT THE BOLUS BUTTON DOES NOT WORK, EVEN AFTER SEVERAL PRESSES. THE KEYPAD IS REPORTEDLY INTACT . THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 5 YR