COULTER® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2011-00841
- Event Type
- Malfunction
- Date Received
- July 6, 2011
- Date of Event
- June 7, 2011
- Report Date
- June 7, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
QC RUN PRIOR TO THE EVENT RECOVERED WITHIN RANGE. THE INSTRUMENT IS CURRENTLY PERFORMING WITHIN QC SPECIFICATIONS. A FIELD SERVICE ENGINEER (FSE) VERIFIED THAT THE APERTURE VOLTAGES AND FLOW RATE WERE SET CORRECTLY. THE FSE VERIFIED THE INSTRUMENT, REVIEWED THE LASER OFFSET VOLTAGES, AND VALIDATED THE SYSTEM. RAW DATA ANALYSIS INDICATES: PATIENT 1: THE PLT COUNT (~222) IS IN THE NORMAL RANGE AND THE OVERALL PATTERN ON THE SPECIMEN RUN IS QUITE NORMAL. FROM A PATTERN PERSPECTIVE. PLATELET CLUMPS ARE NOT OBSERVED FROM THE INSTRUMENT RESULTS. PATIENT 2: INTERFERENCE IS DETECTED ON THE FRONT OF THE WBC HISTOGRAM BUT NOT SUFFICIENT TO SUSPECT THAT THE PLATELET RESULTS ARE NOT CORRECT AND NOT SUFFICIENT TO GENERATE A SUSPECT MESSAGE. PER LABELING, THE FOLLOWING ARE LISTED AS KNOW INTERFERING SUBSTANCES FOR THE PLATELET PARAMETER: GIANT PLATELETS, PLATELET CLUMPS, WHITE CELL FRAGMENTS, ELECTRONIC NOISE, VERY SMALL RED CELLS, RED CELL FRAGMENTS. ROOT CAUSE PER RAW DATA ANALYSIS IS UNKNOWN.
A CUSTOMER CONTACTED BECKMAN COULTER INC., (BEC) AND REPORTED THAT PLATELET (PLT) COUNTS ON 2 SPECIMENS ANALYZED ON THE COULTER LH 750 HEMATOLOGY ANALYZER WERE NOT ACCURATE BECAUSE CLUMPED PLTS WERE OBSERVED ON A MANUAL REVIEW OF THE SMEARS AND STATED THE LH750 IS NOT GENERATING A PLT CLUMPS FLAG WHEN PLT CLUMPS ARE PRESENT ON SMEARS. THE CUSTOMER DID NOT REPORT THE PLT COUNTS OUTSIDE OF THE LABORATORY, BUT INDICATED THE PLTS WERE "CLUMPED BUT ADEQUATE". THERE WAS NO EFFECT TO PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER® LH 750 HEMATOLOGY ANALYZER | AUTOMATED DIFFERENTIAL CELL COUNTER | GKZ | BECKMAN COULTER, INC. | NA | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |