FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 2151731 · Received July 6, 2011

Report

Report Number
2124215-2011-06196
Event Type
Injury
Date Received
July 6, 2011
Date of Event
April 5, 2011
Report Date
April 5, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME, THIS LEAD REMAINS IMPLANTED AND IN SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A PATIENT FOLLOW UP, THE PATIENT WITH THIS RIGHT VENTRICULAR (RV) LEAD REPORTED FEELING CHEST PAIN. AN ECHOCARDIOGRAM WAS PERFORMED AND IT WAS DISCOVERED THAT THE PATIENT HAD A PERICARDIAL EFFUSION. IT WAS BELIEVED THAT THE EFFUSION WAS CAUSED BY A PERFORATION OF THE HEART WALL CLOSE TO THE APEX. THE PATIENT WAS TREATED FOR THE EFFUSION AND RESPONDED WELL. HOWEVER, IT WAS NOTED THAT THE PACING IMPEDANCE AND SENSING MEASUREMENTS HAD DECREASED AND THERE WAS COMPLETE LOSS OF CAPTURE AT THE MAXIMUM OUTPUTS. A REVISION WILL BE PERFORMED TO REPOSITION THE LEAD. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE NVY CPI - DEL CARIBE 0296

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L