FDA Adverse Event
Malfunction
Summary report: N
ALTRUA
MDR report key: 2151708
·
Received July 6, 2011
Report
- Report Number
- 2124215-2011-06233
- Event Type
- Malfunction
- Date Received
- July 6, 2011
- Date of Event
- April 4, 2011
- Report Date
- September 25, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
AT THIS TIME THE PRODUCT REMAINS IN SERVICE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED AS NECESSARY.
Additional Manufacturer Narrative · 1
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. THE DEVICE WAS THEN SUBJECTED TO A SERIES OF AUTOMATED DIAGNOSTIC TESTS THAT VERIFY THE PERFORMANCE OF THE PACING, SENSING AND RECORDING FUNCTIONS OF THE DEVICE. THE DEVICE PERFORMED NORMALLY THROUGHOUT ALL TESTS.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WHICH IS PROGRAMMED AT HIGH OUTPUTS, MIGHT BE DEPLETING PREMATURELY.
Description of Event or Problem · 1
ADDITIONAL INFORMATION WAS RECEIVED THAT THIS DEVICE WAS EXPLANTED AND RETURNED FOR RELIABILITY ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALTRUA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | S403 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 302 MO | 5568| S403| MISMATCH| 5071| 1297 |