FDA Adverse Event Injury Summary report: N

CONTAK RENEWAL

MDR report key: 2151690 · Received July 6, 2011

Report

Report Number
2124215-2011-08998
Event Type
Injury
Date Received
July 6, 2011
Date of Event
April 4, 2011
Report Date
April 4, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. ALL LOW VOLTAGE POWER SUPPLY MEASUREMENTS WERE CURRENTLY WITHIN AN APPROPRIATE RANGE; HOWEVER, FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D), WHICH IS PART OF THE SHORTENED REPLACEMENT WINDOW ADVISORY (B)(6) 2009 POPULATION PRODUCT ADVISORY INITIALLY COMMUNICATED ON (B)(6) 2007, HAD A MONITORING VOLTAGE MEASUREMENT OF 2.52 VOLTS AND 13.3 SECONDS CHARGE TIME MEASUREMENT. THE DEVICE HAD BEEN IMPLANTED FOR GREATER THAN 40 MONTHS BUT NOW THERE IS AN ALLEGATION OF PREMATURE BATTERY DEPLETION AGAINST THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H179

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention H179| 4542| MISMATCH| 4517| 0185