FDA Adverse Event Injury Summary report: N

FLEXTEND

MDR report key: 2151688 · Received July 6, 2011

Report

Report Number
2124215-2011-06825
Event Type
Injury
Date Received
July 6, 2011
Date of Event
April 4, 2011
Report Date
May 10, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
PMA / PMN Number
P960006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

TECHNICAL SERVICES DISCUSSED A POSSIBLE LEAD DISLODGEMENT OR LOOSE CONNECTION. THE LOCAL AREA SALES REPRESENTATIVE WAS CONTACTED FOR ADDITIONAL INFORMATION. AT THIS TIME NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT WILL BE UPDATED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

RECORDS INDICATE THIS PRODUCT REMAINS IN SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED FROM THE FIELD INDICATING AN INVASIVE PROCEDURE WAS PERFORMED TO ASSESS THE LEAD. THERE NO SIGNS OF ANY BREAKS IN LEAD INTEGRITY AND TESTING VIA A PACING SYSTEM ANALYZER (PSA) FOUND ADEQUATE MEASUREMENTS. THE LEAD WAS REPOSITIONED TO A MORE SEPTAL POSITION AND AGAIN ADEQUATE MEASUREMENTS WERE OBSERVED. AT THIS TIME, IT WAS SUSPECTED THAT THE HIGH MEASUREMENTS WERE DUE TO POSSIBLE LEAD LOCATION OR PACING VECTOR. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER AND RIGHT VENTRICULAR (RV) PACING LEAD EXHIBITED LOSS OF CAPTURE. AN RV LEAD SAFETY SWITCH WAS ALSO OBSERVED, HOWEVER IMPEDANCE MEASUREMENTS FOLLOWING THE LEAD SAFETY SWITCH WERE 500 OHMS IN BIPOLAR AND 780 OHMS IN UNIPOLAR. THE RV AMPLITUDE MEASUREMENTS WERE 9MV IN BIPOLAR AND 2MV IN UNIPOLAR. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXTEND IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4088

Patients

Seq Age Sex Outcome Treatment
1 79 YR Required Intervention (B)(4)| (B)(4)| (B)(4)