FDA Adverse Event Malfunction Summary report: N

COULTER AC-T DIFF 2 ANALYZER

MDR report key: 2151516 · Received June 24, 2011

Report

Report Number
1061932-2011-00719
Event Type
Malfunction
Date Received
June 24, 2011
Date of Event
January 4, 2008
Report Date
January 4, 2008
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K990352
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2008, THE FIELD SERVICE ENGINEER REPLACED THE WBC BATH AND PROBE ASSEMBLY AND VERIFIED INSTRUMENT OPERATIONS. THE OPERATOR WAS IN THE PROCESS OF CLEANING THE BATHS AND REPLACING THE WBC CHECK VALVES WHEN SHE ACCIDENTALLY CLOSED THE CAP PIERCE DOOR AND PIERCED HER FINGER. THE OPERATOR DID NOT FOLLOW THE SERVICE AND MAINTENANCE PROCEDURES, REPLACING CHECK VALVES AND CLEANING THE CLOSED VIAL STATION LOCATED IN THE OPERATOR'S GUIDE INSTRUCTING HER TO TURN INSTRUMENT OFF PRIOR TO REPLACING CHECK VALVES. ROOT CAUSE WAS USER ERROR. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW CONDUCTED BETWEEN (B)(4), 2008 TO (B)(4), 2010 OF COMPLAINTS FOR ADD'L REPORTABLE EVENTS.

Description of Event or Problem · 1

CUSTOMER REPORTED A MINOR INJURY SUSTAINED WHILE OPERATING THE COULTER AC-T DIFF 2 ANALYZER. WHILE TROUBLESHOOTING THE ANALYZER WITHOUT ASSISTANCE FROM BECKMAN COULTER INC PERSONNEL, THE OPERATOR ACCIDENTALLY PIERCED HER FINGER ON THE PROBE. THE OPERATOR WAS REPLACING THE PNEUMATIC CHECK VALVES ATTACHED TO THE WHITE BLOOD CELL (WBC) BATH AND SHE ACTIVATED THE ANALYZER RESULTING IN THE PROBE PIERCING HER FINGER. THE ANALYZER WAS POWERED ON AT THE TIME OF THE EVENT. THE OPERATOR WAS WEARING APPROPRIATE PERSONAL PROTECTIVE EQUIPMENT. THE OPERATOR WAS TREATED BY A PHYSICIAN AND FIRST AID WAS ADMINISTERED. HER FINGER WAS BANDAGED. THE OPERATOR HAS SINCE TESTED NEGATIVE FOR INFECTIOUS DISEASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER AC-T DIFF 2 ANALYZER GKZ BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK