FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 2151463 · Received July 6, 2011

Report

Report Number
2124215-2011-06247
Event Type
Injury
Date Received
July 6, 2011
Date of Event
April 4, 2011
Report Date
April 4, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE LEAD WAS BEING RETURNED FOR ANLAYSIS. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS ATRIAL LEAD WAS EXPLANTED AS A RESULT OF IMPEDANCES GREATER THAN 1800 OHMS AND LOSS OF CAPTURE. FLUOROSCOPY ALSO NOTED A LEAD FRACTURE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4469

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention 6987| 4469| 4087| 1294