FDA Adverse Event Malfunction Summary report: N

TRYPTICASE SOY BROTH

MDR report key: 21513983 · Received March 4, 2025

Report

Report Number
MW5167173
Event Type
Malfunction
Date Received
March 4, 2025
Date of Event
February 10, 2025
Report Date
February 27, 2025
Manufacturer
REMEL, INC.
Product Code
JSG
Product Problem
Yes
Report Source
Voluntary report
Reporter Occupation
RISK MANAGER
Health Professional
Yes

Narratives

Description of Event or Problem · 0

THIS AFFECTED MULTIPLE PATIENTS (86) WHO HAD TISSUE SPECIMEN CULTURES. ONE OF THE MICRO COLLEAGUES REPORTED SEEING A TREND WITH GRAM STAIN RESULTS OF STREPLIKE GRAM-POSITIVE COCCI IN CHAINS IN MULTIPLE SPECIMENS. NONE OF THE CULTURES WERE GROWING ANY ORGANISMS. AFTER AN EXTENSIVE INVESTIGATION, IT WAS DETERMINED THAT THE TRYPTICASE SOY BROTH THAT WE WERE USING TO GRIND THE TISSUE IN WAS CONTAMINATED. TEN BOTTLES OF THE BROTH WAS GRAM STAINED AND STREPLIKE GRAM-POSITIVE COCCI WERE FOUND IN THE TSB(TRYPTICASE SOY BROTH) BROTH. REFERENCE REPORTS: MW5167166, MW5167167, MW5167168, MW5167169, MW5167170, MW5167171, MW5167172, MW5167174, MW5167175.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2631554 TRYPTICASE SOY BROTH CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL JSG REMEL, INC. 185090

Patients

Seq Age Sex Outcome Treatment
1 67 YR Male