FDA Adverse Event Malfunction Summary report: N

LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 2151389 · Received June 8, 2011

Report

Report Number
3015876-2011-00456
Event Type
Malfunction
Date Received
June 8, 2011
Date of Event
April 19, 2011
Report Date
May 12, 2011
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K991910
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED BIPHASIC PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A FAILURE OF AN INTEGRATED CIRCUIT CHIP DESIGNATOR U9.

Description of Event or Problem · 1

IT WAS INITIALLY REPORTED THAT THE DEVICE HAD ITS SERVICE INDICATOR ILLUMINATED AND HAD LOGGED A FAULT CODE. PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. THE BIPHASIC PCB ASSEMBLY WAS REPLACED AND PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED BIPHASIC PCB ASSEMBLY AND OBSERVED THAT THE DEFIBRILLATION FUNCTIONALITY WAS DISABLED. THERE WAS NO PT USE ASSOCIATED WITH THE REPORTED FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA