Summary report: N
LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 2151389
·
Received June 8, 2011
Report
- Report Number
- 3015876-2011-00456
- Event Type
- Malfunction
- Date Received
- June 8, 2011
- Date of Event
- April 19, 2011
- Report Date
- May 12, 2011
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- MKJ
- PMA / PMN Number
- K991910
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED BIPHASIC PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A FAILURE OF AN INTEGRATED CIRCUIT CHIP DESIGNATOR U9.
Description of Event or Problem · 1
IT WAS INITIALLY REPORTED THAT THE DEVICE HAD ITS SERVICE INDICATOR ILLUMINATED AND HAD LOGGED A FAULT CODE. PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. THE BIPHASIC PCB ASSEMBLY WAS REPLACED AND PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED BIPHASIC PCB ASSEMBLY AND OBSERVED THAT THE DEFIBRILLATION FUNCTIONALITY WAS DISABLED. THERE WAS NO PT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES | MKJ | PHYSIO-CONTROL, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |