CONTAK RENEWAL
Report
- Report Number
- 2124215-2011-07782
- Event Type
- Injury
- Date Received
- July 6, 2011
- Date of Event
- April 4, 2011
- Report Date
- April 4, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT IN OUR LABORATORY, DEVICE INTERROGATION REVEALED THAT THE DEVICE HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). A LONGEVITY CALCULATION USING THE MOST RECENT PROGRAMMED PARAMETERS AND THERAPY USE INFORMATION INDICATED THAT THE DEVICE FELL SHORT OF LABELED LONGEVITY EXPECTATIONS. REVIEW OF SELF-DIAGNOSTIC TEST RESULTS STORED IN DEVICE MEMORY DID NOT IDENTIFY ANY PERIODS OF EXCESSIVE CURRENT OR ENERGY USE DURING THE LIFE OF THE DEVICE. PRELIMINARY TESTING OF DEVICE FUNCTION DEMONSTRATED NORMAL BEHAVIOR. FURTHER ANALYSIS OF DEVICE MEMORY INDICATED THAT AN ERI INDICATOR WAS DECLARED DUE TO TWO CONSECUTIVE CHARGES EXCEEDING THE ERI CHARGE TIME LIMIT. THE MONITORING VOLTAGE WAS ABOVE THE ERI LIMIT, INDICATING THAT THE BATTERY WAS NOT DEPLETED AND HAD SUFFICIENT CAPACITY REMAINING TO PROVIDE THERAPY IF NEEDED. HOWEVER, THE CHARGE TIMES WOULD HAVE BEEN LONGER THAN THE 18 SECONDS ERI CHARGE TIME LIMIT SPECIFIED IN DEVICE LABELING. DEVICE CIRCUITRY WAS DESIGNED SUCH THAT THE ERI CHARGE TIME LIMIT OF 18 SECONDS WOULD BE REACHED NEAR THE CORRESPONDING ERI MONITORING VOLTAGE LIMIT. IN THIS CASE, A NORMAL BUT EARLIER THAN EXPECTED BUILDUP OF INTERNAL BATTERY IMPEDANCE INCREASED CHARGE TIME, CAUSING THE CHARGE TIME INDICATOR TO TRIP ERI EARLY. DESPITE THE LONGEVITY SHORTFALL, REPLACEMENT INDICATORS OPERATED AS DESIGNED AND THE DEVICE WAS CAPABLE OF DETECTING AND TREATING ARRHYTHMIAS FOR SEVERAL ADDITIONAL MONTHS AS DESCRIBED IN DEVICE LABELING. THIS ENSURED PATIENT SAFETY DURING THE REPLACEMENT PERIOD.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO BATTERY STATUS AT ELECTIVE REPLACEMENT INDICATOR (ERI). THERE WAS A CONCERN THAT THE BATTERY HAD DEPLETED EARLIER THAN EXPECTED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H155 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |