FDA Adverse Event Malfunction Summary report: N

VECTRIS SURESCAN

MDR report key: 21513590 · Received March 4, 2025

Report

Report Number
2649622-2025-06142
Event Type
Malfunction
Date Received
March 4, 2025
Date of Event
February 27, 2025
Report Date
May 19, 2025
Manufacturer
MPRI
Product Code
LGW
PMA / PMN Number
P840001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

G2: THE COUNTRY OF ORIGIN IS CHINA. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: THE LEAD, MODEL# 977A275, SERIAL# (B)(6), WAS RETURNED FOR PRODUCT ANALYSIS. THE RETURNED DEVICE WAS SUBJECTED TO A SERIES OF STANDARD TESTS THAT INCLUDE BUT IS NOT LIMITED TO VISUAL INSPECTION AND ELECTRICAL TESTING. ANALYSIS IDENTIFIED THAT THERE WAS A PROCEDURAL NON-CONFORMANCE; THE DISTAL END OF THE LEAD WAS BENT. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM A DISTRIBUTOR REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT AFTER OPENING THE LEAD PACKAGING, IT WAS FOUND THAT THE LEAD TIP WAS BENT. THE LEAD WAS NOT IMPLANTED AND WAS REPLACED TO COMPLETE THE OPERATION; THE ISSUE WAS RESOLVED. IT WAS NOTED THAT SURGICAL INTERVENTION DID NOT OCCUR AND WAS NOT PLANNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
697621 VECTRIS SURESCAN STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR LGW MPRI 977A275

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown