FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 2151343 · Received June 7, 2011

Report

Report Number
1824206-2011-03108
Event Type
Malfunction
Date Received
June 7, 2011
Date of Event
May 17, 2011
Report Date
May 17, 2011
Manufacturer
HILL-ROM INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THAT THE GAS SPRINGS WERE FROZEN AND COULD NOT BE MOVED. HE REPLACED THE GAS SPRINGS AND TRENDELENBURG FUNCTIONED PROPERLY.

Description of Event or Problem · 1

INFO RECEIVED INDICATES, THE BED COULD NOT BE PLACED INTO TRENDELENBURG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED BIRTHING BED HDD HILL-ROM INC. 3600

Patients

Seq Age Sex Outcome Treatment
1