FDA Adverse Event Injury Summary report: N

CLAVICLE PIN ASSEMBLY 2.5MM

MDR report key: 2151329 · Received June 30, 2011

Report

Report Number
1818910-2011-11839
Event Type
Injury
Date Received
June 30, 2011
Date of Event
June 3, 2011
Report Date
June 3, 2011
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
JDW
PMA / PMN Number
K103001
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE AND X-RAYS ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS FOUND NO MFG DEVIATIONS OR ANOMALIES. A SEARCH OF THE COMPLAINT DATABASE FOUND NO PRIOR REPORTS FOR THIS PART AND LOT NUMBER COMBINATION. A PREVIOUS INVESTIGATION FOUND PRODUCT DEVELOPMENT, MARKETING, AND SURGEON INPUT HAS DETERMINED THAT THE 2.5MM PIN NEEDS TO HAVE A SHORT VERSION FOR SMALL PTS TO HELP PREVENT BREAKAGES. (B)(4) WAS INITIATED ON JUL 28TH, 2009 TO ADDRESS THE CORRECTIVE ACTIONS. (B)(4) WAS COMPLETED ON JUN 10, 2011. THE NEW PART NUMBER FOR THE SHORTER VERSION PIN IS (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ANY ADDITIONAL INFO BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Description of Event or Problem · 1

THE CLAVICLE PIN BROKE DURING INSERTION. THE SURGERY WAS EXTENDED APPROX 30-45 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CLAVICLE PIN ASSEMBLY 2.5MM 87JDW JDW DEPUY ORTHOPAEDICS, INC. NA 101793

Patients

Seq Age Sex Outcome Treatment
1 17 YR