FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2151293 · Received June 7, 2011

Report

Report Number
1824206-2011-03114
Event Type
Malfunction
Date Received
June 7, 2011
Date of Event
May 17, 2011
Report Date
May 17, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN INSPECTED BED AND FOUND THE HEAD SECTION HAD VERY SLOW DRIFT DOWN. REPLACED THE HEAD DOWN SOLENOID VALVE, RETESTED THE BED TO RESOLVE THIS ISSUE. BED FOUND IN STORAGE AREA.

Description of Event or Problem · 1

INFO RECEIVED INDICATED THAT THE HEAD SECTION OF THE BED HAS VERY SLOW DRIFT DOWN. NO PT IMPACT. BED FOUND IN STORAGE ROOM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1900

Patients

Seq Age Sex Outcome Treatment
1