FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 2151293
·
Received June 7, 2011
Report
- Report Number
- 1824206-2011-03114
- Event Type
- Malfunction
- Date Received
- June 7, 2011
- Date of Event
- May 17, 2011
- Report Date
- May 17, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN INSPECTED BED AND FOUND THE HEAD SECTION HAD VERY SLOW DRIFT DOWN. REPLACED THE HEAD DOWN SOLENOID VALVE, RETESTED THE BED TO RESOLVE THIS ISSUE. BED FOUND IN STORAGE AREA.
Description of Event or Problem · 1
INFO RECEIVED INDICATED THAT THE HEAD SECTION OF THE BED HAS VERY SLOW DRIFT DOWN. NO PT IMPACT. BED FOUND IN STORAGE ROOM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | P1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |