HOMEPUMP ECLIPSE, 100ML, 100ML
Report
- Report Number
- 2026095-2011-00167
- Event Type
- Malfunction
- Date Received
- June 21, 2011
- Date of Event
- April 11, 2011
- Report Date
- April 11, 2011
- Manufacturer
- I-FLOW CORPORATION
- Product Code
- MEB
- PMA / PMN Number
- K052117
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHARMACIST
Narratives
A REVIEW OF THE DEVICE HISTORY RECORDS (DHR) WAS CONDUCTED FOR THE LOT NUMBER AND INCIDENT REPORTED. THE DEVICE PASSED ALL MANUFACTURING SPECIFICATIONS PRIOR TO RELEASE. INFO PROVIDED FOR THIS INVESTIGATION REVEALED THAT BOTH PUMPS WERE USED PRIOR TO THE 4-HOUR WAITING PERIOD. THE DIRECTIONS FOR USE (DFU) (1303907, REV. D) CONTAIN A WARNING: "THE HOMEPUMP ECLIPSE MUST BE FILLED 4 HOURS PRIOR TO ADMINISTRATION TO INFUSE AT THE LABELED FLOW RATE. IF THE HOMEPUMP ECLIPSE IS USED IMMEDIATELY AFTER FILLING, FLOW RATE MAY INCREASE BY UP TO 50%." THE DFU ALSO STATES: "CAUTIONS: ACTUAL INFUSION TIMES MAY VARY DUE TO THE FOLLOWING: FILLING THE PUMP LESS THAN NOMINAL RESULTS IN FASTER FLOW RATE." RECEIVED TWO EMPTY, USED PUMPS FOR EVALUATION AND INVESTIGATION. THE VISUAL INSPECTION OF PUMP 1 OBSERVED THE CLAMP CLOSED. THE PUMP WAS FILLED WITH NORMAL SALINE SOLUTION TO THE NOMINAL FILL VOLUME OF 100ML. WHEN THE PINCH CLAMP WAS OPENED, THE PUMP INFUSED. A FLOW RATE ACCURACY TEST WAS PERFORMED. PUMP 1 INFUSED WITHIN SPECIFICATION. THE VISUAL INSPECTION OF PUMP 2 OBSERVED THE CLAMP CLOSED. THE PUMP WAS FILLED WITH NORMAL SALINE SOLUTION TO THE NOMINAL FILL VOLUME OF 100ML. WHEN THE PINCH CLAMP WAS OPENED, THE PUMP INFUSED. A FLOW RATE ACCURACY TEST WAS PERFORMED. PUMP 2 INFUSED WITHIN SPECIFICATION. CUSTOMER COMPLAINT WAS NOT CONFIRMED.
DRUG/DILUENT: STERILE WATER. FILL VOLUME: 100ML AND 60ML. FLOW RATE: 100ML/HR. PROCEDURE: UNKNOWN. CATHPLACE: UNKNOWN. FAST FLOW. FOLLOWING A COMPLAINT FROM A PT, PHARMACY TESTED TWO PUMPS FOR FLOW RATE, ONE FILLED TO 100ML AND ONE FILLED TO 60ML. AFTER FILLING, PHARMACY DID NOT WAIT THE 4 HOURS PRIOR TO TEST. THE 100ML PUMP INFUSED IN 37 MINUTES AND THE 60ML PUMP INFUSED IN LESS THAN 30 MINUTES. NO PT CONTACT. NO ADVERSE EVENT OCCURRED. THE PT DISCARDED THE PUMP. PT SUBSEQUENTLY COMPLETED ANTIBIOTIC THERAPY AND HAS BEEN DISCHARGED FROM THE HOME CARE SERVICE. PER DFU: NOMINAL FLOW RATE: 100 ML/HR. NOMINAL FILL VOLUME: 100ML. MAXIMUM FILL VOLUME: 125ML. THE INFUSION DELIVERY TIME IS 1 HOUR WHEN FILLED TO NOMINAL VOLUME AND FLOW RATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMEPUMP ECLIPSE, 100ML, 100ML | ELASTOMERIC PUMP | MEB | I-FLOW CORPORATION | E101000 | 0A2597 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |