FDA Adverse Event Injury Summary report: N

PINNACLE MTL INS NEUT36IDX52OD

MDR report key: 2151209 · Received June 30, 2011

Report

Report Number
1818910-2011-11708
Event Type
Injury
Date Received
June 30, 2011
Date of Event
May 31, 2011
Report Date
May 31, 2011
Manufacturer
DEPUY INTERNATIONAL, LTD.
Product Code
KWA
PMA / PMN Number
K003523
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. THE CUP POSITION REPORTED OF 62 DEGREES IS MORE VERTICAL THAN RECOMMENDED BY SURGICAL TECHNIQUE. OVERLY VERTICAL CUP POSITION CAN CONTRIBUTE TO INCREASED MATERIAL WEAR. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATIONS SINCE THEIR RELEASE FOR DISTRIBUTION INVOLVING ANY OTHER PATIENT. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PT WAS REVISED TO ADDRESS SUBLUXATION AND SUSPECTED INFECTION. ALL CULTURES CAME BACK NEGATIVE, BUT SURGEON DRAINED 50CCS OF A MILKY, MURKY FLUID FROM BOTH INSIDE AND OUTSIDE THE CAPSULE. THE METAL HEAD AND METAL LINER SHOWED SIGNIFICANT SIGNS OF WEAR. THE CUP WAS AT APPROX 62 DEGREES OF INCLINATION AND SOMEWHAT RETROVERTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PINNACLE MTL INS NEUT36IDX52OD 87KWA KWA DEPUY INTERNATIONAL, LTD. NA 2721741

Patients

Seq Age Sex Outcome Treatment
1 NA Required Intervention