FDA Adverse Event Malfunction Summary report: N

VIPER2 SET SCREW INSERTER, SELF-RETAINING

MDR report key: 2151194 · Received June 21, 2011

Report

Report Number
1526439-2011-00108
Event Type
Malfunction
Date Received
June 21, 2011
Date of Event
May 19, 2011
Manufacturer
DEPUY SPINE, INC.
Product Code
LXH
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION OF THE INSERTERS FOUND TIP BREAKAGE OCCURRED AT THE BASE OF THE FLUTES ON EACH INSTRUMENT. REVIEW OF THE DEVICE HISTORY RECORD FOUND THE LOT OF 266 UNITS MET SPECIFICATION REQUIREMENTS WHEN RELEASED TO STOCK. NO DEFINITIVE CONCLUSIONS CAN BE MADE AS TO THE CAUSE OF TIP BREAKAGE. HOWEVER, THE DAMAGE IS INDICATIVE OF THE APPLICATION OF ATYPICAL FORCE DURING SET SCREW TIGHTENING AND/OR MATERIAL FATIGUE FROM USAGE OVER TIME. THE INSTRUMENTS HAVE BEEN IN SERVICE FOR APPROXIMATELY THREE YEARS. AT THIS TIME, THE COMPLAINT IS CONSIDERED TO BE CLOSED.

Description of Event or Problem · 1

INTERNATIONAL AFFILIATE REPORTS THAT THE TIPS OF TWO VIPER2 SET SCREW INSERTERS, BOTH FROM LOT X0508, BROKE OFF WHILE INSERTING SET SCREWS DURING SPINAL SURGERY. THE BROKEN TIPS BECAME EMBEDDED WITHIN THE SET SCREWS AND COULD NOT BE REMOVED AS THE SET SCREWS WERE LOCKED IN PLACE. THERE WAS NO DELAY TO SURGERY AS ANOTHER INSERTER WAS AVAILABLE. IT WAS REPORTED THAT THE SURGEON IS CONFIDENT THAT NO ADVERSE OUTCOMES WILL OCCUR. AS UNINTENDED PORTIONS OF THE INSERTERS REMAIN IN THE PATIENT, THIS MEDWATCH REPORT IS BEING SUBMITTED TO DOCUMENT THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIPER2 SET SCREW INSERTER, SELF-RETAINING INSTRUMENT, ORTHOPAEDIC, MANUAL, SURGICAL LXH DEPUY SPINE, INC. NA X0508

Patients

Seq Age Sex Outcome Treatment
1 UNK