FDA Adverse Event Malfunction Summary report: N

LAP-BAND ACCESS PORT KIT (UNK SIZE)

MDR report key: 2151163 · Received June 21, 2011

Report

Report Number
2024601-2011-00504
Event Type
Malfunction
Date Received
June 21, 2011
Date of Event
September 27, 2010
Report Date
May 24, 2011
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TAPER II. MEDWATCH SENT TO FDA ON: (B)(4) 2011. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS NOTED THE ACCESS PORT HAD NON-PENETRATING NICKS WITH STRIATIONS DESCRIBED AS SURGICAL TOOL SCRATCH-LIKE. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."

Description of Event or Problem · 1

HEALTH PROFESSIONAL REPORTED LAP-BAND SYSTEM EXPLANT DUE TO AN ALLEGED PORT LEAK. THE DEVICE HAS BEEN REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND ACCESS PORT KIT (UNK SIZE) LTI ALLERGAN NA NI

Patients

Seq Age Sex Outcome Treatment
1 52 YR