FDA Adverse Event Malfunction Summary report: N

BD INSYTE AUTOGUARD

MDR report key: 21509717 · Received March 3, 2025

Report

Report Number
1710034-2025-00349
Event Type
Malfunction
Date Received
March 3, 2025
Date of Event
February 17, 2025
Report Date
April 30, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903814442
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

D. POTENTIAL LOT (BATCH) #S PROVIDED: OUR BEST GUESS IS EITHER LOT # 4222173 OR 4310980 BASED ON UNUSED PRODUCT IN STORAGE. H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: OUR QUALITY ENGINEER INSPECTED THE SAMPLE SUBMITTED FOR EVALUATION. YOUR REPORT OF A LEAK WAS CONFIRMED, AND THE CAUSE WAS ATTRIBUTABLE TO A CIRCUMFERENTIAL CRACK IN THE GREEN CATHETER ADAPTER, WHICH APPEARED TO BE MANUFACTURING RELATED. THE ADAPTER WAS INTACT AND IN ONE PIECE; HOWEVER, THE ADAPTER WAS BROKEN, AND THE PLANE OF THE BREAK WAS ADJACENT TO THE PUSH TAB.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE AUTOGUARD CRACK IN HUB WITH BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN THE IV CATHETER WAS INSERTED INTO THE PATIENT IT STARTED LEAKING BLOOD AROUND THE GREEN HUB PART OF THE CATHETER. IT WAS DISCOVERED THAT THE GREEN HUB HAD A CRACK THROUGH IT. CUSTOMER RESPONSE ON (B)(6) 2025. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT? 02-17-25 DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCURRED AS A RESULT OF THE EVENT. PATIENT REQUIRED A SECOND VENIPUNCTURE TO BE PERFORMED DUE TO FAILURE OF THE INITIAL CATHETER.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1147353 BD INSYTE AUTOGUARD PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. UNKNOWN 00382903814442

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown