BD INSYTE AUTOGUARD
Report
- Report Number
- 1710034-2025-00349
- Event Type
- Malfunction
- Date Received
- March 3, 2025
- Date of Event
- February 17, 2025
- Report Date
- April 30, 2025
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- UDI-DI
- 00382903814442
- PMA / PMN Number
- K201075
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- 003
Narratives
D. POTENTIAL LOT (BATCH) #S PROVIDED: OUR BEST GUESS IS EITHER LOT # 4222173 OR 4310980 BASED ON UNUSED PRODUCT IN STORAGE. H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
INVESTIGATION RESULTS: OUR QUALITY ENGINEER INSPECTED THE SAMPLE SUBMITTED FOR EVALUATION. YOUR REPORT OF A LEAK WAS CONFIRMED, AND THE CAUSE WAS ATTRIBUTABLE TO A CIRCUMFERENTIAL CRACK IN THE GREEN CATHETER ADAPTER, WHICH APPEARED TO BE MANUFACTURING RELATED. THE ADAPTER WAS INTACT AND IN ONE PIECE; HOWEVER, THE ADAPTER WAS BROKEN, AND THE PLANE OF THE BREAK WAS ADJACENT TO THE PUSH TAB.
IT WAS REPORTED THAT BD INSYTE AUTOGUARD CRACK IN HUB WITH BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN THE IV CATHETER WAS INSERTED INTO THE PATIENT IT STARTED LEAKING BLOOD AROUND THE GREEN HUB PART OF THE CATHETER. IT WAS DISCOVERED THAT THE GREEN HUB HAD A CRACK THROUGH IT. CUSTOMER RESPONSE ON (B)(6) 2025. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT? 02-17-25 DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCURRED AS A RESULT OF THE EVENT. PATIENT REQUIRED A SECOND VENIPUNCTURE TO BE PERFORMED DUE TO FAILURE OF THE INITIAL CATHETER.
NO ADDITIONAL INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1147353 | BD INSYTE AUTOGUARD | PERIPHERAL IV CATHETERS | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | UNKNOWN | 00382903814442 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |