FDA Adverse Event
Malfunction
Summary report: N
9600
MDR report key: 2150929
·
Received June 21, 2011
Report
- Report Number
- 1720753-2011-08344
- Event Type
- Malfunction
- Date Received
- June 21, 2011
- Date of Event
- March 18, 2011
- Report Date
- June 21, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP REPLACED THE INTERCONNECT CABLE, THE X-RAY CONTROL BUTTON AND THE CENTRAL PROCESSING BATTERY. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT RETAIN THE DATE AND TIME, AND INTERMITTENTLY WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |