HIP HEAD,10/12 TAPER,28MM,+0MM
Report
- Report Number
- 1818910-2011-11603
- Event Type
- Injury
- Date Received
- June 29, 2011
- Date of Event
- March 16, 2007
- Report Date
- May 31, 2011
- Manufacturer
- DEPUY INTERNATIONAL, LTD.
- Product Code
- KWY
- PMA / PMN Number
- K893872
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
DEPUY HAS RECEIVED 108 ULTIMA TPS FEMORAL STEM EXPLANTS FROM THE (B)(6) HOSPITAL. THESE CONSIST LARGELY OF COMPONENTS FROM A CLINICAL STUDY. THE EXPLANTED STEMS HAVE DISPLAYED SOME EVIDENCE OF CORROSION ON THE SURFACE OF THE IMPLANT. THIS HAS BEEN REPORTED BY THE KEY SURGEONS FROM (B)(6) IN A CLINICAL PAPER (REFERENCE JBJS VOL. 92-B, NO. 11, NOVEMBER 2010). FOLLOWING PREVIOUS CLINICAL REVIEW IN (B)(4) 2007, DEPUY HAD PROVIDED INFORMATION TO USERS REGARDING OUR FINDINGS AND OFFERED GUIDANCE ON THE FUTURE CLINICAL MANAGEMENT OF PATIENTS. ON (B)(4) 2007, THE MHRA ISSUED A MEDICAL DEVICE ALERT TO USERS ((B)(4)). THE ULTIMA TPS HIPS SYSTEM IS NO LONGER AVAILABLE ON THE MARKET FOR CLINICAL USE. MOST OF THE EXPLANTS RECENTLY RECEIVED ARE FAILURES REPORTED IN CLINICAL PAPER AND WHICH LEAD TO THE 2007 MDA [1] (19 OF WHICH HAD PREVIOUSLY BEEN INVESTIGATED THOUGH THE COMPLAINTS SYSTEM). THEY WERE COLLECTED IN (B)(6) AND PASSED TO US FOR ANALYSIS IN 2011. A FEW ARE FROM FURTHER REVISIONS CARRIED OUT AFTER THE MDA FROM THE ORIGINAL COHORT OF PATIENTS. A HEALTH HAZARD EVALUATION ((B)(4)) IS ONGOING AT THIS TIME.
NECROSIS OF PERI-IMPLANT BONE AND SOFT TISSUE FROM MASSIVE METALLOSIS. CAUSED PAIN. MRI POSITIVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HIP HEAD,10/12 TAPER,28MM,+0MM | 87KWY | KWY | DEPUY INTERNATIONAL, LTD. | NA | 1106373 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Required Intervention |