FDA Adverse Event Malfunction Summary report: N

ENDOPATH XCEL TROCAR

MDR report key: 2150834 · Received July 6, 2011

Report

Report Number
3005075853-2011-02784
Event Type
Malfunction
Date Received
July 6, 2011
Date of Event
June 14, 2011
Report Date
June 16, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD CONDITION. UPON EVALUATION OF THE DEVICE, THE DUCKBILL WAS FOUND TO BE SLIGHTLY OPEN AT SLIT. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK WITHOUT THE TEST PROBE INSERTED THROUGH THE DEVICE. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. HOWEVER, AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE INSUFFLATIONS ISSUES. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRECTOMY PROCEDURE, ABDOMINAL AIR LEAKED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.