FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2150823 · Received July 6, 2011

Report

Report Number
2531779-2011-04705
Event Type
Malfunction
Date Received
July 6, 2011
Date of Event
May 23, 2011
Report Date
June 8, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE DOWN ARROW BUTTON IS INTERMITTENTLY RESPONDING TO PRESSES. THE PATIENT CONFIRMED THAT THE KEYPAD IS INTACT AND THE BUTTONS ARE NOT STICKING. THE PATIENT REPORTEDLY WORE THE PUMP IN A POCKET AND DID NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 18 YR