FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2150729 · Received July 6, 2011

Report

Report Number
2531779-2011-04701
Event Type
Malfunction
Date Received
July 6, 2011
Report Date
June 10, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/29/2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE DOWN-ARROW AND OK BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER THE KEY CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE DOWN ARROW BUTTON DOES NOT RESPOND. THE PATIENT DENIES DAMAGE TO KEYPAD OR EXPOSURE TO WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 37 YR