ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2150721 · Received July 6, 2011

Report

Report Number
2531779-2011-04699
Event Type
Malfunction
Date Received
July 6, 2011
Report Date
June 10, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE UP-ARROW AND DOWN-ARROW BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THE KEYPAD WAS REMOVED AND ADHESIVE WAS FOUND UNDER THE CONTACTS. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE. (B)(6).

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT THE UP ARROW BUTTON REQUIRES SEVERAL PRESSES FOR A RESPONSE. SHE DENIED HOLES/TEARS IN THE KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 13 YR