FDA Adverse Event
Malfunction
Summary report: N
ALLEGRETTO WAVE EYE-Q
MDR report key: 2150672
·
Received June 22, 2011
Report
- Report Number
- 3003288808-2011-00192
- Event Type
- Malfunction
- Date Received
- June 22, 2011
- Report Date
- May 23, 2011
- Manufacturer
- WAVELIGHT GMBH
- Product Code
- LZS
- PMA / PMN Number
- P030008/P0
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4). THIS REPORT WAS MAILED TO FDA ON: (B)(4) 2011. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTS THE EYE TRACKER PRESENTED DIFFICULTY IN FINDING THE PUPIL. DURING FOLLOW-UP, THE CUSTOMER CONFIRMED BSS (BALANCED SALT SOLUTION) SPATTER ON THE LASER OUTPUT LENSES. IT IS SUSPECTED THIS OCCURRED ON THE EYE TRACKER. NO PT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALLEGRETTO WAVE EYE-Q | OPHTHALMIC EXCIMER LASER SYSTEM | LZS | WAVELIGHT GMBH | 8065990631 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |