FDA Adverse Event
Malfunction
Summary report: N
KNIFE
MDR report key: 2150658
·
Received June 22, 2011
Report
- Report Number
- 2523835-2011-00093
- Event Type
- Malfunction
- Date Received
- June 22, 2011
- Report Date
- May 23, 2011
- Manufacturer
- ALCON PRECISION DEVICE - SINKING SPRING
- Product Code
- HNN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHARMACIST
Narratives
Additional Manufacturer Narrative · 1
THE INVESTIGATION IS IN PROGRESS. SAMPLES HAVE NOT YET BEEN RECEIVED FOR EVALUATION. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT WAS REVIEWED. FUNCTIONAL PENETRATION TEST VALUES FOR THE LOT MET SPECIFICATION. NO ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO A DULL KNIFE WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO COMPANY'S ACCEPTANCE CRITERIA. THE ROOT CAUSE HAS NOT BEEN IDENTIFIED. (B)(4).
Description of Event or Problem · 1
A PHARMACIST REPORTED DURING SURGERY, THE KNIFE WAS NOTED TO HAVE HAD POOR CUTTING PERFORMANCE. A SECOND KNIFE WAS USED AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KNIFE | MANUAL OPHTHALMIC SURGICAL INSTRUMENT | HNN | ALCON PRECISION DEVICE - SINKING SPRING | 8065982265 | 831017M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |