COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-10477
- Event Type
- Malfunction
- Date Received
- July 6, 2011
- Date of Event
- June 19, 2011
- Report Date
- June 22, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K041191
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE CUSTOMER'S REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH THE SCREEN HAVING HORIZONTAL REPEATING LINES WAS CONFIRMED BUT NOT REPRODUCED. THE CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE A DAMAGED MAIN DISPLAY. THE MAIN DISPLAY WAS REPLACED TO FIX THE REPORTED CONDITION. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP THAT THE SCREEN HAS HORIZONTAL REPEATING LINES. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THE EVENT OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION NECESSARY. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |