FDA Adverse Event Malfunction Summary report: N

TERUMO VIRTUOSAPH VEIN HARVESTING SYSTEM

MDR report key: 2150614 · Received June 22, 2011

Report

Report Number
1124841-2011-00270
Event Type
Malfunction
Date Received
June 22, 2011
Date of Event
June 6, 2011
Report Date
June 6, 2011
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
GCJ
PMA / PMN Number
K031891
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

TERUMO DID RECEIVE THE ACTUAL DEVICE; VISUAL INSPECTION NOTED MINOR SCRATCHES, DENTS ON THE ROD AND EYEPIECE OF THE ENDOSCOPE. THIS TYPE OF DAMAGE IS OFTEN ATTRIBUTED TO HANDLING ISSUES. THE DEVICE IFU STATES THAT THE ENDOSCOPE MUST BE HANDLED WITH EXTREME CARE TO PREVENT DAMAGE TO THE DEVICE. (B)(4). THE DEVICE IFU STATES TO CHECK THE IMAGE QUALITY BEFORE AND AFTER EACH USE TO CHECK FOR SIGNS OF DAMAGE. IF IMAGE IS UNACCEPTABLE, CONTACT TERUMO CVS CUSTOMER SERVICE FOR ASSISTANCE. ALL AVAILABLE INFORMATION HAS BEEN PLACED ON FILE IN QUALITY MANAGEMENT FOR APPROPRIATE TRACKING, TRENDING AND FOLLOW UP.

Description of Event or Problem · 1

THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR SYSTEMS THAT DURING VEIN HARVESTING PROCEDURE, THE ENDOSCOPE WAS CLOUDY. THE PRODUCT WAS CHANGED OUT. THERE WAS NO HARM TO THE PATIENT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO VIRTUOSAPH VEIN HARVESTING SYSTEM ENDOSCOPE GCJ TERUMO CARDIOVASCULAR SYSTEMS CORP. MCENDO550R NA

Patients

Seq Age Sex Outcome Treatment
1 UNK