TERUMO VIRTUOSAPH VEIN HARVESTING SYSTEM
Report
- Report Number
- 1124841-2011-00270
- Event Type
- Malfunction
- Date Received
- June 22, 2011
- Date of Event
- June 6, 2011
- Report Date
- June 6, 2011
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- GCJ
- PMA / PMN Number
- K031891
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
TERUMO DID RECEIVE THE ACTUAL DEVICE; VISUAL INSPECTION NOTED MINOR SCRATCHES, DENTS ON THE ROD AND EYEPIECE OF THE ENDOSCOPE. THIS TYPE OF DAMAGE IS OFTEN ATTRIBUTED TO HANDLING ISSUES. THE DEVICE IFU STATES THAT THE ENDOSCOPE MUST BE HANDLED WITH EXTREME CARE TO PREVENT DAMAGE TO THE DEVICE. (B)(4). THE DEVICE IFU STATES TO CHECK THE IMAGE QUALITY BEFORE AND AFTER EACH USE TO CHECK FOR SIGNS OF DAMAGE. IF IMAGE IS UNACCEPTABLE, CONTACT TERUMO CVS CUSTOMER SERVICE FOR ASSISTANCE. ALL AVAILABLE INFORMATION HAS BEEN PLACED ON FILE IN QUALITY MANAGEMENT FOR APPROPRIATE TRACKING, TRENDING AND FOLLOW UP.
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR SYSTEMS THAT DURING VEIN HARVESTING PROCEDURE, THE ENDOSCOPE WAS CLOUDY. THE PRODUCT WAS CHANGED OUT. THERE WAS NO HARM TO THE PATIENT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TERUMO VIRTUOSAPH VEIN HARVESTING SYSTEM | ENDOSCOPE | GCJ | TERUMO CARDIOVASCULAR SYSTEMS CORP. | MCENDO550R | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |