FDA Adverse Event
Malfunction
Summary report: N
TORNIER
MDR report key: 2150548
·
Received June 22, 2011
Report
- Report Number
- 9610667-2011-00014
- Event Type
- Malfunction
- Date Received
- June 22, 2011
- Date of Event
- May 25, 2011
- Report Date
- May 25, 2011
- Manufacturer
- TORNIER INC.
- Product Code
- HXX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
IT IS REPORTED THAT DURING THE IMPLANTATION OF A SHOULDER PROSTHESIS THAT A DISTAL SEGMENT 3.5MM HEX TIP SCREW DRIVER TIP BROKE OFF WHILE TIGHTENING THE SCREW OF THE GLENOSPHERE COMPONENT. THE TIP SEGMENT COULD NOT BE EXTRACTED FROM THE SCREW HEAD SOCKET. THIS IS A KNOWN POTENTIAL ISSUE OF FAILURE WHERE EXCESSIVE TORQUE HAS BEEN APPLIED IN TOOL USE. THE SCREWDRIVER IS A CLASS 1 SURGICAL TOOL. THIS REPORT IS PROVIDED TO IDENTIFY AN EVENT OF INADVERTENT RESIDUAL FOREIGN BODY LEFT INDWELLING AT THE SURGICAL SITE WITHIN THE IMPLANT DEVICE. REMAINING SEGMENT OF THE TOOL WILL BE AVAILABLE FOR EXAMINATION. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TORNIER | SCREW DRIVER, 3.5MM, HEX TIP | HXX | TORNIER INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |