AUTOCAT2 WAVE
Report
- Report Number
- 1219856-2011-00214
- Event Type
- Malfunction
- Date Received
- June 22, 2011
- Date of Event
- June 8, 2011
- Report Date
- June 20, 2011
- Manufacturer
- ARROW INTL., INC.
- Product Code
- DSP
- PMA / PMN Number
- K060309
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE WILL NOT BE RETURNED FOR EVAL.
IT WAS REPORTED VIA A CALL REPORT AT 0011 FROM THE NURSE IN THE CVICU (CARDIOVASCULAR INTENSIVE CARE UNIT) THAT THE CONDENSATION BOTTLE HAD BEEN EMPTIED NUMEROUS TIMES AND THE ALARM WAS STILL OCCURRING EVERY 60-90 MINS. THE PT CURRENTLY IS IN SINUS RHYTHM RATE IN THE 90S. THEY ARE PLANNING TO REMOVE THE INTRA-AORTIC BALLOON (IAB) IN THE AM. IT WAS EXPLAINED BY THE CLINICAL SUPPORT SPECIALIST (CSS) THAT SHE DID NOT HAVE TO ATTEMPT TO EMPTY THE BOTTLE WITH EVERY ALARM. THE PUMP MAY NOT HAVE ENOUGH TIME TO COMPLETE THE DRAIN TASK WITH THE DIASTOLIC TIME OR A VALVE COULD BE STICKING. IT WAS EXPLAINED HOW TO PERFORM A MANUAL PURGE AND WHEN TO COMPLETE THIS TASK. SHE WAS ASKED TO CALL THE CSS BACK IF THE ALARM BEGINS TO OCCUR EVERY 20 MINS AS THE PUMP MAY NEED TO BE CHECKED OUT BY BIOMED. AT 0122AM THE SECOND CALL WAS RECEIVED WITH THE ALARM OCCURRING EVERY 20 MINS AND THE PT'S HEART RATE WAS CONSISTENTLY IN THE HIGH 80S TO LOW 90S. IT WAS SUGGESTED THAT THE INTRA-AORTIC BALLOON PUMP (IABP) BE CHANGED OUT FOR ANOTHER AND THIS BE SENT TO BIOMED SO THE VALVES COULD BE ASSESSED. THE NURSE WAS COMFORTABLE CHANGING THE IABP FOR ANOTHER. THERE WAS NO PT DEATH, INJURIES OR COMPLICATIONS. IT IS UNK IF THERAPY WAS DELAYED WHILE THE PUMP WAS EXCHANGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOCAT2 WAVE | INTRA-AORTIC BALLOON PUMP PRODUCTS | DSP | ARROW INTL., INC. | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |