FDA Adverse Event
Injury
Summary report: N
PHACO PLUS
MDR report key: 21505
·
Received April 27, 1995
Report
- Report Number
- MW1005747
- Event Type
- Injury
- Date Received
- April 27, 1995
- Date of Event
- February 1, 1995
- Report Date
- March 21, 1995
- Manufacturer
- ALLERGAN MEDICAL OPTICS
- Product Code
- HQC
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- WA, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
PHACO MACHINE WAS PRIMED; ONCE PRIMED ORIGINAL SETTINGS ARE USUALLY LOW VACUUM. IT WAS NOT DOUBLE CHECKED, AS INITIAL SCULPTING WAS PERFORMED, THE SURGEON NOTICED A PIECE OF IRIS PULLED INTO PHACO AND ASKED FOR READING OF SETTINGS. IT WAS THEN NOTED THE HIGH VACUUM SETTING WAS LIGHTED ON THE PANEL. COMPUTER BOARD ON MACHINE REPLACED ON 3/1/95.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PHACO PLUS | PHACOEMULSIFICATION UNIT | HQC | ALLERGAN MEDICAL OPTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention |