OMNIPOD 5 PODS
Report
- Report Number
- 3004464228-2025-08669
- Event Type
- Malfunction
- Date Received
- March 3, 2025
- Date of Event
- February 2, 2025
- Report Date
- March 3, 2025
- Manufacturer
- INSULET CORPORATION
- Product Code
- QFG
- UDI-DI
- 20385083000579
- PMA / PMN Number
- K231826
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HAS NOT BEEN RETURNED/RECEIVED TO DATE. IF THE DEVICE IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED WITH THE INVESTIGATION RESULTS. IT WAS REPORTED THAT THE CANNULA HAD DISLODGED FROM THE INFUSION SITE. THIS CONDITION COULD INTERRUPT INSULIN DELIVERY AND CONTRIBUTE TO HYPERGLYCEMIA. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING COMMUNICATING WITH THE PDM, DEPLOYMENT, DELIVERING FLUID, OCCLUSION DETECTION, AND FREEDOM FROM HAZARD ALARMS. CLOUD - LOCKED DOWN/SMARTPHONE PHONE_CONTROL_ANDROID. CLOUD - OMNIPOD 5 SOFTWARE APP VERSION 3.1.1. CLOUD - SMARTPHONE OPERATING SYSTEM TQ3A.230901. 001.C2. CLOUD - SMARTPHONE HARDWARE PIXEL 7. CLOUD - CGM SENSOR TYPE G6. PLEASE NOTE, THAT SECTION D IS CAPTURING THE DEVICE IDENTIFIERS AS REPORTED BY THE COMPLAINANT. THIS MAY NOT ALIGN TO THE DEVICE CONFIGURATION REPORTED IN H11, AS THIS DATA IS PULLED FROM OUR CLOUD BASED ON THE REPORTED DATE OF EVENT.
IT WAS REPORTED THE PATIENT'S BLOOD GLUCOSE LEVELS (BG) ROSE TO OVER 250 MG/DL, WHILE WEARING THE POD BETWEEN 4 AND 24 HOURS. THE POD REPORTEDLY FELL OFF THE INFUSION SITE (ARM), CAUSING THE CANNULA TO DISLODGE. AS TREATMENT, THE PATIENT CHANGED OUT THE POD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 634595 | OMNIPOD 5 PODS | OMNIPOD 5 PODS | QFG | INSULET CORPORATION | PT-001662 | PH1K05242431 | 20385083000579 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | Female |