FDA Adverse Event Malfunction Summary report: N

MICRO SAGITTAL SAW

MDR report key: 2150298 · Received June 7, 2011

Report

Report Number
1811755-2011-02054
Event Type
Malfunction
Date Received
June 7, 2011
Date of Event
May 10, 2011
Report Date
May 10, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

INTERNAL CORROSION IN THE MOTOR AND ROTOR ASSEMBLY WAS IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE.

Description of Event or Problem · 1

THE MICRO SAGITTAL SAW WAS SENT IN FOR EVALUATION DUE TO OVERHEATING. THERE WAS NO PATIENT INVOLVEMENT, NO ADVERSE EVENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO SAGITTAL SAW HBE STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK