FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2150030 · Received July 5, 2011

Report

Report Number
6000001-2011-10292
Event Type
Malfunction
Date Received
July 5, 2011
Date of Event
June 7, 2011
Report Date
June 7, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE A FAULTY USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB). TO CORRECT THIS CONDITION, THE UIM PCB WAS REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A COLLEAGUE INFUSION PUMP REQUIRED THE REPLACEMENT OF THE TRIPLE CHANNEL USER INTERFACE PRINTED CIRCUIT BOARD ASSEMBLY (PCBA). THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1