FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM
MDR report key: 2150006
·
Received July 5, 2011
Report
- Report Number
- 2050012-2011-02835
- Event Type
- Malfunction
- Date Received
- July 5, 2011
- Date of Event
- June 6, 2011
- Report Date
- June 6, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WAS ON-SITE (B)(4)2011. FSE FOUND THAT THE Y FITTING WAS BROKEN ON ONE CONNECTION PORT AND REPLACED THE Y FITTING AND ALL TUBING ON THE NO FOAM BOTTLE. THE INSTRUMENT WAS PRIMED 5 TIMES WITH NO FURTHER LEAKING OBSERVED. REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REPORTING A NO FOAM LEAK IN THE UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEM. THE NO FOAM TUBING CAME OFF OF THE Y CONNECTOR FITTING. THE CUSTOMER REATTACHED THE TUBING AND IT NO LONGER APPEARED TO BE LEAKING. THE OPERATOR SLIPPED ON THE WET FLOOR BUT WAS NOT INJURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |