FDA Adverse Event Injury Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2149090 · Received May 26, 2011

Report

Report Number
1423500-2011-06557
Event Type
Injury
Date Received
May 26, 2011
Date of Event
July 12, 2008
Report Date
July 12, 2008
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS MDR IS BEING SUBMITTED AS PART OF RETROSPECTIVE SUMMARY REPORT # (B)(4). THE TOTAL NUMBER OF EVENTS BEING SUMMARIZED ON THIS MDR ARE 101 - "FKX REPORTABLE DEATH AND SERIOUS INJURIES". THESE REPORTS ARE BEING FILED LATE TO FULFILL BAXTER'S COMMITMENT TO PERFORM A TWO YEAR RETROSPECTIVE REVIEW OF RENAL COMPLAINTS IN RESPONSE TO FDA-WARNING - LETTER (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED A BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) REGARDING A SYSTEM ERROR 2240 ALARM THAT APPEARED ON THE DISPLAY OF THE HOME PATIENT'S HOMECHOICE MACHINE DURING THEIR AUTOMATED PERITONEAL DIALYSIS THERAPY. PER THE INITIAL REPORT, THE PATIENT LINE WAS NOT CONNECTED TO THE TRANSFER SET BEFORE PRESSING GO TO BEGIN INITIAL DRAIN/THERAPY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT. ADDITIONAL INFORMATION WAS OBTAINED FROM THE HOME PATIENT'S NURSE ON (B)(6) 2008 REGARDING A REPORT OF PERITONITIS AND CLOUDY EFFLUENT. THE PATIENT WAS ADMITTED TO THE HOSPITAL ON (B)(6) 2008. THE NURSE BELIEVES THE CAUSE OF THE PERITONITIS WAS DUE TO CONTAMINATION BASED ON THE CULTURE RESULTS. THE PATIENT WAS TREATED WITH VANCOMYCIN (DOSAGE UNKNOWN). THE NURSE ALSO STATED THE PATIENT WAS NEW TO THE AREA AND HAD UNDERGONE TRAINING WITH THE PREVIOUS CLINIC. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2008, BUT WAS READMITTED TO THE HOSPITAL WITH A (B)(6) INFECTION ON (B)(6) 2008. PER THE NURSE, THE PATIENT IS STILL IN THE HOSPITAL RECOVERING FROM THE (B)(6) INFECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE FKX - SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME NA NI

Patients

Seq Age Sex Outcome Treatment
1 60 YR Hospitalization| R LO/CAL 6L DIANEAL (B)(4) | LO/CAL 6L DIANEAL (B)(4) | LO/CAL 6L DIANEAL (B)(4) | LO/CAL 6L DIANEAL (B)(4)| LO/CAL 6L DIANEAL (B)(4)| LO/CAL 6L DIANEAL (B)(4)