FDA Adverse Event Malfunction Summary report: N

TANGO OPTIMO

MDR report key: 2149028 · Received June 15, 2011

Report

Report Number
9610824-2011-00081
Event Type
Malfunction
Date Received
June 15, 2011
Date of Event
May 17, 2011
Report Date
June 15, 2011
Manufacturer
BIO-RAD MEDICAL DIAGNOSTICS GMBH
Product Code
KSZ
PMA / PMN Number
BK080013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

THE CUSTOMER HAD REPORTED POSITIVE REACTIONS OF FOUR SAMPLES AFTER SOLIDSCREEN II CONTROL WAS PERFORMED. FURTHER TESTINGS OF CUSTOMER CONFIRMED ONE CORRECT POSITIVE SAMPLE AND THREE FALSE POSITIVE SAMPLES. WE RECEIVED NEITHER THE SAMPLES NOR THE AFFECTED LOT OF ANTI-HUMAN GLOBULIN ANTI-IPG SOLIDSCREEN II. THEREFORE THE RETENTION SAMPLES OF THE AFFECTED LOTS WERE TESTED. WE ALSO PROCESSED SOLIDSCREEN II CONTROL AND ANTIBODY FREE SAMPLES. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. WE DIDN'T OBSERVE ANY FALSE POSITIVE REACTIONS AFTER RUNNING THE POSITIVE CONTROL. THE TESTING QUALITY CONTROL LABORATORY CONFIRMED THE CORRECT FUNCTION OF THE AFFECTED LOTS. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE AFFECTED NEGATIVELY THE QUALITY OF THE COMPLAINED LOT. SERVICE INTERVENTION ON TANGO OPTIMO WAS PERFORMED ON (B)(6) 2011. SEVERAL PARTS INCLUDING RIGHT PIPETTOR NEEDLE AND ACTIVE WASH KNF PUMP ON TANGO OPTIMO S/N (B)(4) WERE REPLACED. PERFORMANCE VERIFICATION USING THE SAMPLES IN QUESTION, REVIEW OF TANGO EVENTLOG LIST AND RESULTS REVEALED NO FURTHER OCCURRENCE OF THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TANGO OPTIMO AUTOMATED BLOOD BANK ANALYZER SYSTEM KSZ BIO-RAD MEDICAL DIAGNOSTICS GMBH

Patients

Seq Age Sex Outcome Treatment
1 MLB 2: LOT 7036120, EXP. 8/28/2012| SOLIDSCREEN II STRIP: LOT 7046080, EXP. 11/11/2012| EXP. 2/25/2012| AHG ANTI-IGG SOLIDSCREEN II: LOT 7035091-05,