FDA Adverse Event Malfunction Summary report: N

BUBBLE CPAP CIRCUIT

MDR report key: 2148925 · Received July 5, 2011

Report

Report Number
9611451-2011-00405
Event Type
Malfunction
Date Received
July 5, 2011
Date of Event
June 9, 2011
Report Date
June 10, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
REFER TO H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4), PMA/510(K): THE (B)(4) BUBBLE CPAP CIRCUIT IS NOT SOLD IN THE USA, BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K020332. MANUFACTURER NARRATIVE: THE BC161 BUBBLE CPAP CIRCUIT IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE RECEIVE THE COMPLAINT DEVICE, AND HAVE COMPLETED OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). THE BC161 BUBBLE CPAP CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K020332. MANUFACTURER NARRATIVE THE COMPLAINT BC161 BUBBLE CPAP CIRCUIT WAS NOT RETURNED TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WITHOUT THE COMPLAINT DEVICE, WE ARE UNABLE TO DETERMINE THE ROOT CAUSE OF THE REPORTED FAULT. ALL BREATHING CIRCUITS ARE TESTED FOR CONNECTIVITY AND ELECTRICAL CONTINUITY DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THE CIRCUITS ARE ALSO REJECTED IF THEY ARE UNABLE TO FIT INTO THE TESTING DEVICE. IT IS THEREFORE MOST LIKELY THAT THE PIN OF THE COMPLAINT BC161 CIRCUIT WAS DAMAGED POST PRODUCTION, POSSIBLY DURING THE SET-UP AND CONNECTION OF THE EQUIPMENT.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THE HEATER WIRE PINS OF A NEWLY OPENED BC161 BUBBLE CPAP CIRCUIT KIT WERE BROKEN. THIS WAS NOTICED PRIOR TO PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THE HEATER WIRE PINS OF A NEWLY OPENED BC161 BUBBLE CPAP CIRCUIT KIT WERE BROKEN. THIS WAS NOTICED PRIOR TO PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BUBBLE CPAP CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED BC161 101203