FDA Adverse Event Injury Summary report: N

NEEDLE 27GA 1-1/2IN

MDR report key: 21469379 · Received February 26, 2025

Report

Report Number
1911916-2025-00153
Event Type
Injury
Date Received
February 26, 2025
Date of Event
February 14, 2025
Report Date
March 16, 2025
Manufacturer
BECTON DICKINSON
Product Code
FMI
UDI-DI
30382903016298
PMA / PMN Number
K021475
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP FOR DEVICE EVALUATION IT WAS REPORTED THE NEEDLE SNAPPED AND REMAINED IN THE PATIENT. TO AID IN THE INVESTIGATION, THREE SAMPLES IN SEALED PACKAGING BLISTERS WERE RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED WITH 30X MAGNIFICATION. NO DEFECTS OR IMPERFECTIONS WERE OBSERVED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 301629, LOT 4178893. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. BASED ON THE INVESTIGATION WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED.

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED.

Description of Event or Problem · 0

MATERIAL # 301629. BATCH # 4212478. VERBATIM: (B)(6) RECEIVED A COMPLAINT VIA PHONE. PERFORMING A THYROID BIOPSY THE NEEDLE SNAP AND WAS STUCK IN THE PATIENT¿S NECK AND THE PATIENT HAD TO HAVE SURGERY TO REMOVE IT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1039274 NEEDLE 27GA 1-1/2IN NEEDLE, HYPODERMIC, SINGLE LUMEN FMI BECTON DICKINSON 4178893 30382903016298

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention